Horizontal Vessel With Volume Measurement - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

Parameter
Parameter No. /Name
type
Timer (2.5.2.)
Material Level (2.5.1.)
Independent
Auto False Echo Suppression
parameters
Range (2.8.7.2.)
Auto False Echo Suppression
(2.8.7.1.)
a)
Distance to product from sensor reference point.
b)
For more details see Auto False Echo Suppression Range (2.8.7.2.) on page 91 and Auto
False Echo Suppression (2.8.7.1.) on page 90.
Return to Measurement: press Mode

Horizontal vessel with volume measurement

Notes:
• Minimum distance from flange face to target is limited by Near Range (2.8.1.).
• We recommend using Auto False Echo Suppression with an empty or almost empty
vessel.
To obtain level measurement/4 to 20 mA output proportional to vessel volume in a chemical
vessel:
Low Calibration Point = 3.5 m (11.48 ft) from sensor reference point
High Calibration Point = 0.5 m (1.64 ft) from sensor reference point
Max. fill/empty rate = 0.2 m/min (0.65 ft/min)
Select vessel shape, Parabolic Ends, and enter values for A and L, to obtain a volume reading
instead of level.
In the event of a loss of echo: SITRANS LR200 is to go into Fail-safe High after 2 minutes.
sensor
reference
point
3.5 m
0.5 m
High
Cal
Point
Low Cal
Point
Page 140
SITRANS LR200 (HART) – INSTRUCTION MANUAL
to start normal operation.
A = 0.8 m
Options/Values Function (cont'd)
2
HI
a)
Distance
minus
0.42 m (1.37 ft)
LEARN
L = 6 m
2 minutes
Fail-safe level set to High
Sets length of learned
b)
TVT
to use
Enables the use of learned
TVT
100% = 8000 l
volume
reading
A = 0.8 m
7ML19985JP02