Environmental
Notes:
Approvals
• Check
install.
• Use appropriate conduit seals to maintain IP or NEMA rating.
• location
• altitude
• ambient temperature
• relative humidity
• installation category
• pollution degree
Process
• process temperature
- polypropylene rod
- PTFE rod or SS horn
• pressure (vessel)
Approvals
The device nameplate lists the approvals that apply to your device.
Note:
• General
• Radio
• Hazardous
1)
The maximum temperature is dependent on the process connection, antenna materials,
and vessel pressure. For more detail, or for other configurations, see
Temperature Chart
page 132 onwards.
2)
Intrinsically Safe wiring
See
7ML19985JP02
on page 11, for the specific configuration you are about to use or
indoor/ outdoor
5000 m (16,404 ft) max.
−40 to +80 °C (−40 to +176 °F)
suitable for outdoor
Type 4X/NEMA 4X, Type 6/NEMA 6, IP67, IP68
enclosure (see note below)
I
4
1)
−40 to +80 °C (−40 to +176 °F)
−40 to +200 °C (−40 to +392 °F)
1)
3 bar, gauge (43.5 psi, gauge)
CSA
, FM, CE
US/C
Europe (R&TTE), FCC, Industry Canada
2)
Intrinsically Safe
:
(Europe)
(US/Canada)
Process Pressure/Temperature derating curves
on page 131, and
on page 22.
SITRANS LR200 (HART) – INSTRUCTION MANUAL
ATEX II 1 G, EEx ia IIC T4
FM/CSA: (barrier required)
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Class III T4
Maximum Process
on
Page 11