Parameter
Parameter No./Name
type
Material (1.2.)
Response Rate (1.3.)
Units (1.4.)
Quick Start
Operating Mode (1.5.)
Wizard
parameters
Low Calibration Point (1.6.)
High Calibration Point (1.7.)
Apply? (Apply changes) (1.8.)
Vessel Shape (2.7.1.1.)
Maximum Volume (2.7.1.2.)
Indepen-
Dimension A (2.7.1.3.)
dent
parameters
Dimension L (2.7.1.4.)
Timer (2.5.2.)
Material Level (2.5.1.)
Auto False Echo Suppression
Range (2.8.7.2.)
Auto False Echo Suppression
(2.8.7.1.)
a)
Distance to product from sensor reference point.
b)
For more details see
False Echo Suppression (2.8.7.1.)
Return to Measurement: press Mode
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Options/Values Function
LIQUID
MED
M
LEVEL
3.5
0.5
YES
PARABOLIC ENDS Defines vessel shape.
8000
0.8
6
2
HI
[Distance
0.42 m]
LEARN
Auto False Echo Suppression Range (2.8.7.2.)
on page 90.
to start normal operation.
Medium =1 m/minute
meters
Level is reported as Volume
when a vessel shape is selected.
3.5 m (11.48 ft)
0.5 m (1.64 ft)
Transfers Quick Start settings to
device.
8000 liters
0.8 m (2.62 ft)
6 m (19.68 ft)
2 minutes
Fail-safe level set to High
a)
minus
Sets length of learned TVT
b)
curve
to use
Enables the use of learned TVT
b
curve
.
on page 91 and
Auto
Page 141