Appendix J: Firmware Revision History
Firm-
PDM
ware
EDD
Rev.
Rev.
1.00
1.00
1.01
1.02
1.03
2.00
1.04
2.00
7ML19985JP02
Date
6 May 2003
•
Initial Release
•
Factory calibration of antenna offset can now
26 Jun 2003
be saved.
•
Sporadic Fault 25 (electronic failure) messages
16 Oct 2003
reduced.
•
Cleaned up fault code issues.
16 Mar 2004
•
Set HART status bit 4 (more status available)
•
MEASURE key works only on restricted set of
parameters while in program mode.
•
Made corrections to language strings.
•
Corrected User Interface operation at –30
•
Fixed Fault 25 (electronic failure) caused by an
RTM glitch on the RAMP_TH line.
•
PDM TVT profile was sometimes lacking a few
bytes at the end. This is now fixed.
•
Added TVT Shaper to HART: P831 to cmd 172
and create cmd 174 for P832.
•
Improved mA output time at startup.
•
Product startup time reduced from about 30
seconds to about 10 seconds.
•
P838 now does not reset.
•
P831 and P832 TVT shaper added.
•
P054 and P054 breakpoints corrected for PDM.
•
Fixed possible corruption of HART Status.
27 May 2004
•
Prevent possible buffer overflows.
•
Fixed HART command 17 to correct status
information.
•
Changes made to the program mode timer to
avoid an immediate timeout followed by a
switch to run mode when the device is
switched into program mode via HART.
•
The configuration changed status can now be
saved for the duration of a power cycle.
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Changes
o
C.
Page 173