Environmental Specifications

Specification
Ambient Operating
Temperature
Optimum Operating
Temperature
Relative Humidity
Optimum Relative
Humidity
Indoor Use Only
Altitude
Ventilation
IEC 1010-1 Installation Category II
IEC 1010-1 Equipment Classification Class I
IEC 1010-1 Pollution Degree 2

Electrical Requirements

Requirement
Electrical Rating
Power Requirements
Maximum Power Input
Fuse Rating
Line Leakage Current
124
Value
18°C to 30°C (64°F to 86°F).
At temperatures under 22°C (72°F), urobilinogen
and leukocyte results may be decreased, and at
temperatures above 26°C (79°F), increased.
22°C to 26°C (72°F to 79°F).
20% to 80%, non-condensing, actively controlled
35% to 55%
up to 2000 meters
1709 BTU
Value
100 to 240 VAC ± 10%
50 to 60 Hz
72 VA
2 A, 250 V, 2 AG, SB(T)
<0.5 mA in normal condition
<3.5 mA in single fault condition