Omega Engineering OMB-NETSCAN 1500 User Manual page 162

Ethernet/internet-based data logging & control instrument
Table of Contents

Advertisement

API Command Reference
TYPE
EXECUTION
SYNTAX
DESCRIPTION
When an error has occurred, the ERROR indicator light on the front panel of the unit turns on. Error Status
Query (
the unit responds with one of the following error codes:
: No error has occurred.
E000
E001
: Invalid device dependent command (IDDC) due to a command syntax error.
E002
: Invalid device dependent command option (IDDCO). A command parameter was out of range or
missing.
E004
: A Channel Configuration Error indicates that a channel has been inappropriately configured, either
because the channel is not present or because the specified channel type is not compatible with the card type
installed.
E008
: A Calibration Error occurs when the card calibration has failed or has been inappropriately calibrated.
E016
: A Trigger Overrun Error occurs when more than one trigger event occurs for a given trigger state or when
a non-expectant trigger event occurs.
E032
: An Open T/C Error or Range Error, occurs when the hardware circuitry detects an open thermocouple, or
when the software detects the A/D input has reached or exceeded its linearization.
: A Command Conflict Error indicates the issuance of a command that is in conflict with other commands
E128
that have been issued or the current configuration. This error usually occurs when a command is issued that
cannot be performed because of the current state of the unit. For instance, if the Set Scan Interval (
is issued during an acquisition (
After execution of the Error Status Query, most error conditions are cleared. Error conditions may also be
cleared by issuing a Clear Event Status (
or
command is executed to clear the error condition.
U0
EXAMPLES
PRINT#1,"E?X"
INPUT#1,A$
PRINT A$
PRINT#1,"K3 X"
PRINT#1,"E?X"
INPUT#1,A$
PRINT A$
A-22
E?
System
Immediate
E?
Returns present error condition of the unit with one of the following error
E000
E001
E002
E004
E008
E016
E032
E128
E?
) is used to determine the present error condition. After execution of the Error Status Query command,
tstart 0
- Error Status Query
codes:
- No error has occurred.
- Invalid device dependent command (IDDC).
- Invalid device dependent command option (IDDCO).
- Channel configuration error.
- Calibration error.
- Trigger overrun.
- Open T/C or Range Error.
- Command conflict error.
).
U0
) command to the unit. The ERROR light will remain on until an
' Request error status
' Display shows
E000
' Send illegal command option to the unit
' (Note: ERROR indicator light turns on)
' Request error status
' (Note: ERROR indicator light turns off)
' Displays
(Invalid Device Dependent Command Option)
E002
, no errors occurred
NetScan User's Manual
Appendix A
I
) command
E?

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omb-netscan

Table of Contents