API Command Reference
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Clear Channel Configuration (
configured. After this command has been performed, channel readings will be not be updated; therefore, no HLL
or acquisition buffer updates will be performed until a new configuration is entered via the Configure Channels (
command. This command may be used when the present configuration is no longer acceptable and you wish to
reconfigure channels from scratch.
EXAMPLE
PRINT#1,"*CX"
PRINT#1,"C1, 1X"
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Restore Factory Defaults (
when it left the factory. Issuing this command will cause the unit to load the factory default configuration into the
working configuration. The factory default configuration is defined in the Configuring Power-Up State section of
Appendix D.
This command may not be issued while an acquisition is configured since it will cause certain settings upon
which the acquisition depends to change. Issuing this command while an acquisition is configured will cause a
Conflict Error to occur.
Note: Issuing the
EXAMPLE
PRINT#1,"Q?T?Y?X"
LINE INPUT#1, A$
PRINT A$
PRINT#1,"*FX"
PRINT#1,"Q?T?Y?X"
LINE INPUT#1,A$
PRINT A$
A-12
*C
- Clear Channel Configuration
Channel
Deferred
*C
Clear out the channel configuration so that no channels are configured
*C
*F
- Restore Factory Settings
System
Immediate
*F
Restore unit to factory configuration
*F
) command may be used to restore the unit to the same configuration it had
command will cause any previously defined configuration to be irrevocably lost. The
*F
calibration factors, however, will remain intact.
) command will clear out the channel configuration so that no channels are
' Clear all channels configuration
' Configure channel 1 of T/C type J
' Retrieve some current configuration settings
' Get the settings
' Screen shows
001000,000000
' Restore factory default configuration
' Retrieve some current configuration settings
' Get the settings
' Screen shows
000000,000000
Q1,7,7,7,0 T0,3,0,0 Y000100,
Q1,0,0,0,0 T0,0,0,0 Y000000,
NetScan User's Manual
Appendix A