Omega Engineering OMB-NETSCAN 1500 User Manual page 114

Ethernet/internet-based data logging & control instrument
Table of Contents

Advertisement

2.
Apply calibration voltage from the precision calibrator to the first channel of each CSN14/HV/S card
(see following table), and verify with a digital multimeter. Requirements for both devices are specified
in Required Equipment.
Volts DC
-2.5 V
+2.5 V
-25.0 V
+25.0 V
-250.0 V
+250.0 V
Note: In the slot columns, the digit after the G is the channel number and the number immediately preceding
the X assigns the voltage value, e.g., G1,20X means "Calibrate Gain for channel 1, ±2.5 V."
3.
After each calibration type
Response codes and meanings applicable to calibration are as follows:
Response to
4.
After completing the gain calibration of all cards, type
5.
Type
Note: You can view the calibration constants by typing an entry similar to the following:
C#1X QC?
You can monitor the calibration voltage by typing an entry similar to the following:
*B
C1,1X
T1,1,0,0X;@X
U13X
Note that
6-16 Calibration
Slot # 1
Slot # 2
G1,20X
G17,20 X
G1,20X
G17,20X
G1,21X
G17,21X
G1,21X
G17,21X
G1,22X
G17,22X
G1,22X
G17,22X
U2X
Indication for Calibration Mode
U2X
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature sensor out of range)
U144
Invalid password
U160
Invalid command
to confirm calibration mode is disabled. A return of
U2X
...for slot #1
...to clear the buffer
...to addresses the channel, this example is for slot #1, channel 1
... to trigger the acquisition (Trigger LED flashes); );
every entry of U13X results in an updated value of voltage in counts
can be used to start or stop the acquisition.
@X
Slot # 3
Slot # 4
Slot # 5
G33,20X
G49,20X
G65,20X
G33,20X
G49,20X
G65,20X
G33,21X
G49,21X
G65,21X
G33,21X
G49,21X
G65,21X
G33,22X
G49,22X
G65,22X
G33,22X
G49,22X
G65,22X
and check for a return of
U128
EX
Slot # 6
Slot # 7
G81,20X
G97,20X
G81,20X
G97,20X
G81,21X
G97,21X
G81,21X
G97,21X
G81,22X
G97,22X
G81,22X
G97,22X
to confirm no errors.
to end the calibration mode.
confirms this.
U000
starts acquisition
@X
NetScan User's Manual
Slot # 8
G113,20X
G113,20X
G113,21X
G113,21X
G113,22X
G113,22X

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omb-netscan

Table of Contents