API Command Reference
EXAMPLE
PRINT#1, "F0,0X"
PRINT#1, "*CX"
PRINT#1,"C1,1,-100.0, 100.0, 0.5"
PRINT#1,"C33-34,11"
PRINT#1,"C65,2"
PRINT#1,"X"
PRINT#1,"U13X"
LINE INPUT#1,A$
PRINT A$
TYPE
EXECUTION
SYNTAX
DESCRIPTION
Using the Select Card (
cause subsequent
calibration factors of the cards as part of the system, select a number from the range
EXAMPLE
PRINT#1, "C#0X"
PRINT#1, "QC?X"
LINE INPUT#1, A$
PRINT A$
PRINT#1, "C#5X"
PRINT#1, "U12X"
LINE INPUT#1, A$
PRINT A$
PRINT#1, "C#105X"
PRINT#1, "QC?X"
LINE INPUT#1, A$
PRINT A$
Note: In the above example an expansion chassis is being used, thereby allowing an additional 4 card slots.
Note: Card #5 will typically be located in slot #5; however, if the optional relay card is used, card #5 will be located in
slot #6 and signal conditioning card use is limited to 7 cards (instead of 8).
A-20
C#
- Select Card
Calibration
Immediate
C#card
Select the card from which subsequent
retrieve information.
Without optional Relay Card
Valid options for
card in corresponding slot of the main unit (with 1 being the bottom slot);
and 5, 6, 7, or 8 - Selects the corresponding slot on the expansion chassis
(with slot 5 being the bottom slot of the expansion chassis.
With optional Relay Card Installed (installs in slot 1)
Valid options for
2 selects card in slot 3; and so on, with 7 selecting the card in slot 8; and
no card 8 available.
C#
) command, selecting a number from the range
QC?
commands to return the calibration factors of the cards by themselves. To return the
' Set Data Format to Engineering Units, degrees C
' Clear current channel configuration
' Configure channel for T/C type J with low setpoint of -
100.0 °C, high setpoint of 100.0 °C and hysteresis of
0.5 °C
' Configure channels 33 and 34 for Volts type (100 mV
range) and no setpoints
' Configure channel 65 for T/C type K with no setpoints
' Execute the above configuration (The unit is now
scanning the above configuration)
' Get the last readings of configured channels
' Retrieve response
' Screen shows
+0034.40
-0103.00
+0012.30
+0004.60
QC?
card
are:
0
- Selects the chassis; 1, 2, 3, or 4 - Selects the
card
are:
0
- Selects the chassis; 1 selects card in slot 2,
1-8
' Select chassis
' Read chassis calibration factors
' Select the card #5
' Read back the time/date that card #5 was last calibrated
' Select the card #5 in order to read back the calibration factors
as part of the system
' Read back the system calibration factors of card #5
Appendix A