Omega Engineering OMB-NETSCAN 1500 User Manual page 112

Ethernet/internet-based data logging & control instrument
Table of Contents

Advertisement

2.
Apply calibration voltage from the precision calibrator to the first channel of each card (see following
table), and verify with a digital multimeter. Requirements for both devices are specified in Required
Equipment.
Volts DC
-50 mV
+50 mV
-100 mV
+100 mV
-1.0 mV
+1.0 mV
-5.0 mV
+5.0 mV
-10 mV
+10 mV
Note: In the slot columns, the digit after the G is the channel number and the number immediately preceding
the X assigns the voltage value, e.g., G1,1X means "Calibrate Gain for channel 1, ±50 mV."
3.
After each calibration type
Response codes and meanings applicable to calibration are as follows:
Response to
4.
After completing the gain calibration of all cards, type
5.
Type
Note: You can view the calibration constants by typing an entry similar to the following:
C#1X QC?
You can monitor the calibration voltage by typing an entry similar to the following:
*B
C1,1X
T1,1,0,0X;@X
U13X
Note that
6-14 Calibration
Slot # 1
Slot # 2
G1,1X
G17,1X
G1,1X
G17,1X
G1,11X
G17,11X
G1,11X
G17,11X
G1,12X
G17,12X
G1,12X
G17,12X
G1,13X
G17,13X
G1,13X
G17,13X
G1,14X
G17,14X
G1,14X
G17,14X
U2X
U2X
Indication for Calibration Mode
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature sensor out of range)
U144
Invalid password
U160
Invalid command
to confirm calibration mode is disabled. A return of
U2X
...for slot #1
...to clear the buffer
...to addresses the channel, this example is for slot #1, channel 1
... to trigger the acquisition (Trigger LED flashes); );
every entry of U13X results in an updated value of voltage in counts
can be used to start or stop the acquisition.
@X
Slot # 3
Slot # 4
Slot # 5
G33,1X
G49,1X
G65,1X
G33,1X
G49,1X
G65,1X
G33,11X
G49,11X
G65,11X
G33,11X
G49,11X
G65,11X
G33,12X
G49,12X
G65,12X
G33,12X
G49,12X
G65,12X
G33,13X
G49,13X
G65,13X
G33,13X
G49,13X
G65,13X
G33,14X
G49,14X
G65,14X
G33,14X
G49,14X
G65,14X
and check for a return of
U128
EX
Slot # 6
Slot # 7
G81,1X
G97,1X
G81,1X
G97,1X
G81,11X
G97,11X
G81,11X
G97,11X
G81,12X
G97,12X
G81,12X
G97,12X
G81,13X
G97,13X
G81,13X
G97,13X
G81,14X
G97,14X
G81,14X
G97,14X
to confirm no errors.
to end the calibration mode.
confirms this.
U000
starts acquisition
@X
NetScan User's Manual
Slot # 8
G113,1X
G113,1X
G113,11X
G113,11X
G113,12X
G113,12X
G113,13X
G113,13X
G113,14X
G113,14X