114, 115, 116, and 117
Calibration Information
Step
Function
L
18
DC Volts
19
20
21
22
m
23
AC Millivolts
24
E
25
DC Millivolts
26
G Diode
27
E
28
Capacitance
29
30
LoZ
Capacitance
Set calibrator to standby, reconfigure leads, and program for amps output
A
31
DC Amps
?
32
AC Amps
$
33
DC μamps
$
34
AC μamps
y
35
Temperature
36
37
x
38
39
40
10
Table 5. DMM Performance Tests (cont)
Range
Applied
6.000V
0V
6.000V
5V
60.00V
50V
600.0V
600V
600.0V
-600V
600.0 mV
6 mV, 45 Hz
600.0 mV
600mV, 1 kHz
600.0 mV
10 mV
600.0 mV
600mV
2.000 V
1.9V
1000 nF
Open
9999 μF
900 μF
10.00 μF
9 μF
10.00 A
10A
6.000 A
5.0A, 45 Hz
600.0 μA
600 μADC
600.0 μA
600 μAAC, 45 Hz
Open input
0.0 °C
400 °C
0.5 V, 45 Hz
0.5 V, 0 Hz
[2]
500 V
,
500 Hz
Display Reading
[1]
[1]
114
115
-0.002 to 0.002
4.973 to 5.027
49.73 to 50.27
596.8 to 603.2
-596.8 to -603.2
5.6 to 6.4
587.7 to 612.3
9.7 to 10.3
596.8 to 603.2
N/A
N/A
0 to 2
N/A
881 to 919
N/A
9.87 to
N/A
10.13
4.922 to
N/A
5.078
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
0.2 to 0.8,
AC
N/A
Annunciator
On
0.2 to 0.8,
DC
N/A
Annunciator
On
489.7 to
N/A
510.3
[1]
[1]
116
117
1.881 to 1.919
0 to 2
0 to 2
881 to 919
881 to 919
7.18 to 10.82
9.87 to
N/A
10.13
4.922 to
N/A
5.078
593.8 to
N/A
606.2
590.7 to
N/A
609.3
0PEn
N/A
-1.0 to 1.0
N/A
395.0 to
N/A
405.0
0.2 to 0.8,
0.2 to 0.8,
AC
AC
Annunciator
Annunciator
On
On
0.2 to 0.8,
0.2 to 0.8,
DC
DC
Annunciator
Annunciator
On
On
489.7 to
489.7 to
510.3
510.3