The available reading rooms for your Coronis Uniti are:
CR/DR/ MAMMO
CT/MR/NM
Staff Office
Clinical Viewing Room
Emergency Room
Operating Room
To select a reading room:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Ambient Light menu.
3. Enter the Reading Room submenu.
4. Select one of the available reading rooms and confirm.

5.20 Continuous ALC

About Continuous ALC
Continuous ALC can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 35 to correctly set the display func-
tion.
Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged
ambient light into account.
To select continuous ALC:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Ambient Light menu.
3. Enter the Continuous ALC submenu.
4. Select Enabled/Disabled as desired and confirm.

5.21 Embedded QA

Overview
About Embedded QA
DICOM status report
DICOM compliance check
DICOM calibration
Reset DICOM calibration
K5902079 CORONIS UNITI 21/04/2015
Corresponds to light conditions in diagnostic
reading rooms for computed radiology, digital
radiology or mammography. This setting has the
lowest maximum ambient light.
Corresponds to light conditions in diagnostic
reading rooms for computed tomography, magnetic
resonance or nuclear medicine scans.
Corresponds to light conditions in office rooms.
Corresponds to light conditions in diagnostic
reading rooms for clinical viewing.
Corresponds to light conditions in emergency
rooms.
Corresponds to light conditions in operating rooms.
This setting has the highest maximum ambient light.
5. Advanced operation
37