5 Calibration Procedure
6
START
Press and hold
UNIT key for 6
seconds
Display shows
CAL
Display shows
LoAd
Load weight
(see Note 1.)
Press UNIT key once to confirm.
Display shows calibration weight.
END
Note 1:
Capacity
6kg
15kg
35kg
60kg
150kg
300kg
Calibration
Weight
5kg
10kg
20kg
40kg
100kg
200kg