Features; Calibration Procedure (Cal) - KERN 440...N Service Manual

Electronic precision balances
Table of Contents

Advertisement

1 Features

• Full Tare
• Memory function with indicator
• Stable indicator
• Negative value indication
• Two types of Digital Auto Calibration
• Solder pads to prevent end-user internal calibration
• RS232 interface, printing function
• Custom scale by going through SET manual
Multiple serial interface modes
-
Different communication baud rates selectable
-
Auto off enable / disable
-
Zero tracking enable / disable
-
Multiple calibration weight selectable
-
Backlight selectable
-
Animal weighing
-
• Low battery indicator
• Auto off function (battery mode only)
• AC adaptable
• Display segment test function
• Overload protection
• Optional rechargeable battery
• Pre-Tare function

2 Calibration Procedure (CAL)

1. Turn balance on and allow the unit to acclimatize and stabilize for 2 hours.
2. Select the calibration weight according to the operating manual with [Print],
[Mode/Cal] and [Set/M] keys.
3. Press and hold the [Mode/Cal] key until the display shows a flashing weight reading.
The flashing weight reading indicates the correct weight that must be placed on the
weighing pan.
4. Gently place the correct calibration weight on the center on the weighing pan. With the
display still flashing, press the [Set/M] key.
5. The display will now show [Cal F] and then return to normal operating mode. The
calibration is complete when the display correctly shows the weight that is placed on
the weighing pan while the display has stabilized.
6. Remove the calibration weight and press the [TARE] key to reset the zero point.
In case [CAL E] is displayed instead of [CAL F], this indicates a calibration
procedure error. Turn the scale off and then on and repeat the procedure.