6 Inspection Mode for Technical Maintenance

START
Press ON key without releasing, and
press UNIT key 3 times
Display shows zero count
XXXX. Please see note 1
for the zero count
Press TARE key
Display show
0
Note 1:
Capacity
6kg
15kg
35kg
60kg
150kg
300kg
The total internal count is 32768. Overload was set at 32500 (in production).
<--------------------------------------32768 -------------------------------------------------------------------->
<-----
------
---------------------------32500 ------------------------------------------------------> Overload
<-2000~12000 (zero) + 4000~8000 (span)= 6000~20000 -> Safety range
EOE, EOS...N-SH-e-0921
A
Zero Count
2000~12000
2000~12000
2000~12000
2000~12000
2000~12000
2000~12000
Loading calibration weight
(see note 1)
Display shows span count.
Please see Note 1 for span
count range
Adjust RX1 value. Please see
note 1 for recommended RX1
value
Span Count
Calibration
4000~8000
4000~8000
4000~8000
3200~6400
4000~8000
4000~8000
A
Proceed calibration
procedure as per
chapter 5
Span Control
Weight
5kg
10kg
20kg
40kg
100kg
200kg
(RX1)
80K
62K
50K
25K
60K
30K
7