Features; Calibration Procedure (Cal) - KERN ECB Service Manual

Electronic platform scales
Hide thumbs Also See for KERN ECB:
Table of Contents

Advertisement

1 Features

• Zero weight function
• Negative value indication
• Low battery indicator
• Auto zero
• Display segments test function
• Auto off function in battery mode - enable/disable (only ECB)
• Overload protection
• Overload indication
• Digital span calibration
• Digital internal calibration
• Operated by 6 x AA size alkaline batteries
• Rechargeable battery as option
• Zero tracking - enable/disable (only ECB)
• Backlight - 3 modes (only ECB)

2 Calibration Procedure (CAL)

1. Turn scale on and allow the unit to acclimatize and stabilize for 10 minutes.
2. Press and hold the [Tare] key for approx. 3 seconds. Display will show [CAL] and then
the required calibration weight.
3. Gently place the correct calibration weight on the center on the weighing pan.
4. Wait until the display shows [ F] and then return to normal weighing mode
automatically.
5. Remove the calibration weight and press the [TARE] key to reset the zero point.
In case [ E] is displayed instead of [ F], this indicates a calibration
procedure error. Turn the scale off and then on and repeat the procedure.
ECB/ECE-SH-e-0820
3