10.2 Procedure linear calibration
ADJ-SH-e
1910
Press CAL and the screen displays a zero-point count
value, such as 397
Put a 100g and a 200 g standard weight (class E2) in
the center of the weighing pan respectively, read and
record the count value.
The count value range of the balance is shown below
Zero point
350~450
Exit the count value inspection, by pressing ON/OFF
twice to reboot the balance.
Press ON/OFF to shut down the scale.
25
100g
7800~7950
15250~15550
200g