Microlife WatchBP Office Targe Instruction Manual page 18

Office blood pressure measurement device with built-in blood pressure targets
Table of Contents

Advertisement

Error
Description
«Err 3» No pressure
in the cuff
«Err 5» Abnormal
result
17
Potential cause and
remedy
An adequate pressure
cannot be generated
in the cuff. A leak
may have occurred.
Replace the batteries if
necessary. Repeat the
measurement.
The measuring signals
are inaccurate and no
result can therefore
be displayed. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.
Error
Description
«Err 6
Triple Mode There were too many
Potential cause and
remedy
errors during the
measurement in
Triple mode, making it
impossible to obtain a
final reliable result. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.