3.4 Commands
LEXE?
Example:
LCME?
Example:
SIM918 Precision Current Preamplifier
Execution Error
Query the Last Execution Error code. A query of LEXE? always
clears the error code, so a subsequent LEXE? will return 0. Valid
codes are:
Value
Definition
0
No execution error since last LEXE?
1
Illegal value
2
Wrong token
3
Invalid bit
16
Reference clock inactive
*STB? 12; LEXE?; LEXE?
3
0
The error (3, "Invalid bit") is because *STB? only allows bit-specific
queries of 0–7. The second read of LEXE? returns 0.
Command Error
Query the Last Command Error code. A query of LCME? always
clears the error code, so a subsequent LCME? will return 0. Valid
codes are:
Value
Definition
0
No command error since last LCME?
1
Illegal command
2
Undefined command
3
Illegal query
4
Illegal set
5
Missing parameter(s)
6
Extra parameter(s)
7
Null parameter(s)
8
Parameter bu er overflow
10
Bad integer
11
Bad integer token
12
Bad token value
14
Unknown token
*IDN
LCME?
4
The error (4, "Illegal set") is due to the missing "?".