3.4 Commands
CESR? [i]
Example:
CESE(?) [i,] j
Example:
OLSR? [i]
Example:
OLSE(?) [i,] j
Example:
RCSR? [i]
Example:
RCSE(?) [i,] j
Example:
SIM918 Precision Current Preamplifier
Communication Error Status
Query the Communication Error Status Register [Bit i].
Upon executing a CESR? query, the returned bit(s) of the CESR reg-
ister are cleared.
CESR?
0
Communication Error Status Enable
Set (query) the Communication Error Status Enable register [Bit i]
to j .
CESE?
2
Overload Status
Query the Overload Status Register [Bit i].
Upon executing an OLSR? query, the returned bit(s) of the OLSR reg-
ister are cleared.
OLSR?
3
Overload Status Enable
Set (query) the Overload Status Enable register [Bit i] to j .
OLSE 4
Reference Clock Status
Query the Reference Clock Status Register [Bit i].
Upon executing an RCSR? query, the returned bit(s) of the RCSR reg-
ister are cleared.
RCSR?
7
Reference Clock Status Enable
Set (query) the Reference Clock Status Enable register [Bit i] to j .
RCSE 3,1
3 – 17