Sensor Settings; F.2 Sensor Settings - TESTO 335 Instruction Manual

Flue gas analyser
Table of Contents

Advertisement

F. Configuration
32

F.2 Sensor settings

F.2

Sensor settings

It is possible to set an NO
tion/disconnect). The actual calibration data and the status of the measuring cells can
be displayed. Recalibration can be carried out.
Calling up the function:
Sensor settings
Setting the NO2 addition:
1
.
NO2 addition
Option:
Reset N02 addition to default value:
2
Set the value
Change
Setting sensor protection:
For devices with the "Dilute all sensors" option: If a measuring cell threshold value is
exceeded in slot 2, the gas to sensor 2 is diluted by factor five. If a measuring cell
threshold value is exceeded in slot 3, gas to all sensors is diluted by factor two.
With dilution active, the reading resolution and accuracies will change, see Technical
data. Diluted values are represented inversely.
If the threshold is still exceeded despite dilution, the instrument is switched off. To de-
activate sensor protection, set the thresholds to 0 ppm.
addition and thresholds for activating sensor protection (dilu-
2
.
OK
.
OK
.
Deflt
To extend the measuring range and
protect the measuring cells against over-
loads, you can set thresholds which, when
exceeded, activate sensor protection.
Thresholds for a variety of parameters can
be set, depending on the measuring cells
that are connected.
For instruments without "Dilution of all sen-
sors" option: If a threshold of the measur-
ing cell in slot 2 is exceeded, the gas to
sensor 2 is diluted by a factor of five.
There is switch-off if a measuring cell
threshold value is exceeded in slot 3.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents