TESTO 335 Instruction Manual page 52

Flue gas analyser
Table of Contents

Advertisement

K. Technical data
52
K.2 Measuring ranges and accuracies
Parameter
Measuring range
Pabs
600...1150hPa
Temperature (NiCrNi) -40...1200°C
Temperature (PtRhPt)-40...1600°C
Efficiency
0...120%
Flue gas loss
-20.0...99.9%
1
Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
2
Whichever is the greater
With activated dilution sensor 2
Parameter
Measuring range
CO
0...50000ppm
COlow
0...2500ppm
NO2
0...2500ppm
SO2
0...25000ppm
With activated dilution of all sensors (optional)
Parameter
Measuring range
O2
0...25Vol.%
CO, H2-komp.
0...20000ppm
COlow
0...1000ppm
NO2
0...1000ppm
SO2
0...10000ppm
NOlow
0...600ppm
NO
0...6000ppm
1
Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
Accuracy
±10hPa
±0.5°C
at 0.0...100.0°C
±0.5% of reading in rest of range
±1°C
at 0.0...200.0°C
±0.5% of reading in rest of range
-
-
Accuracy
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
Accuracy
±1Vol.% of reading additional error (0...4,99Vol.%) 0.01Vol.%
±0,5Vol.% of reading additional error(5...25Vol.%)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
+10% of reading (additional error)
Resolution
1hPa
0.1°C at -40.0...999.9°C
1°C
at 1000°C...1200°C
1°C
0.1%
0.1%
Resolution
1ppm
0.1ppm
0.1ppm
1ppm
Resolution
1ppm
0.1ppm
0.1ppm
1ppm
0.1ppm
1ppm
t90
1
-
depends
on probe
depends
on probe
-
-
t90
1
t90
1
<20s