Technical Data; Standards And Tests; Measuring Ranges And Accuracies - TESTO 335 Instruction Manual

Flue gas analyser
Table of Contents

Advertisement

K. Technical data
K.1 Standards and tests
· As declared in the certificate of conformity, this product complies with Directive
89/336/EEC.
· This product is TÜV approved to EN 50379 part 2, exception: SO2 and NO2 parame-
ters are not tested, recalibration is not blocke d.
K.2 Measuring ranges and accuracies
Parameter
Measuring range
O2
0...25Vol.%
CO, H2-comp.
0...10000ppm
COlow
0...500ppm
NO2
0...500ppm
SO2
0...5000ppm
NOlow
0...300ppm
NO
0...3000ppm
Draught, ∆p1
-40...40hPa
∆p2
-200...200hPa
Accuracy
±0.2Vol.%
±10ppm or
±10% of reading
2
at 0...200ppm
±20ppm or
±5% of reading
2
at 201...2000ppm
±10% of reading at 2001...10000ppm
±2ppm
at 0.0...39.9ppm
±5% of reading at 40.0...500ppm
±10ppm
at 0.0...200.0ppm
±5% of reading at 200.1...500.0ppm
±10ppm
at 0...200ppm
±5% of reading at 201...2000ppm
±10% of reading at 2001...5000ppm
±2ppm
at 0.0...39.9ppm
±5% of reading at 40.0...300.0ppm
±5ppm
at 0...100ppm
±5% of reading at 101...2000ppm
±10% of reading at 2001...3000ppm
+1.5% v. Mw.
at -40.00...-3.00hPa
+ 0.03hPa
at -2.99...2.99hPa
+ 1.5% v. Mw.
at 3.00...40.00hPa
±1.5% of reading at -200.0...-50.0hPa
±0.5hPa
at -49.9...49.9hPa
±1.5% of reading at 50.0...200.0hPa
K. Technical data
K.1 Standards and tests
Resolution
0.01Vol.%
1ppm
0.1ppm
0.1ppm
1ppm
0.1ppm
1ppm
0.01hPa
0.1hPa
51
1
t90
<20s
<40s
<40s
<40s
<40s
<30s
<30s
-
-