2105 OPERATOR'S MANUAL – SECTION 5
Use the following general guidelines to select the 
best combination of settings to fit your requirements:
OUTPUT
Continuous
High 
Voltage
16, 8 ohm
Mid-Level
4, 2 ohm
High 
Current
1, 0.5 ohm
Also see the "9 Specifications" on page
43section for more information.
5.3.2
Accessing the Power Supply
Board
To access the Power Supply Board, follow the in-
structions given at the beginning of this Section to 
open the Configuration Access Panel. The Power 
Supply Board is located to the right of the Main 
Board, as shown in Figure 5.9.
Figure 5.9 – Accessing the Power Supply Board
5.3.3
Changing Amplifier Voltage
Potential
The 2105 can be configured for High Current (90V) 
or High Voltage (180V) operation via user-
selectable plugs on the Power Supply Board. 
Complete the following steps to change the ampli-
fier voltage potential.
97-2105006_4-2-15
LOAD
Pulse
Settings
16, 8, 4
180 V
ohm
Auto
90 V
2, 1 ohm
Auto
0.75 - 0.25
90 V
ohm
Low
Information subject to change
Figure 5.10 – Location of Amplifier High-Voltage 
Transformer Sockets
Figure 5.11 – J3 and J7 Plug Locations for 
High-Current Output
Figure 5.12 – J4 and J8 Plug Locations for 
High-Voltage Output
1. Locate the two XFMR sections (left of Line 
Voltage sections) (see Figure 5.10).
2. For High Current (90V) Output (see Figure
5.11).      
3. For High Voltage (180V) Output (see Figure
5.12).
        
21