Figure 4.5 - Main Status Indicators For Stand-Alone Amplifiers; Figure 4.4 - Main And Fault Status Indicators - AE Techron 2105 Operator's Manual

Gradient amplifier
Hide thumbs Also See for 2105:
Table of Contents

Advertisement

Enable button is pressed on each amplifier, that 
amplifier will be placed in Ready mode (Ready and 
Standby LEDs lit) and remain in Ready mode until 
all amplifiers in the system have been Enabled. 
The system will automatically proceed to Run
mode when all amplifiers in the system achieve 
Ready mode.
Stop – For stand-alone amplifiers, Stop will place 
the amplifier in Standby mode (both Standby and 
Stop LEDs will be lit). When the amplifier is part of 
a multi-amp system, pressing the Stop button on 
any amplifier in the system will place all amplifiers 
in Standby mode. When an amplifier is in Standby 
mode, the low-voltage transformer is energized but 
the high-voltage transformers are not. 
Reset – For stand-alone amplifiers, when the am-
plifier has stopped due to a fault condition, press-
ing the Reset button will return the amplifier to Run 
mode if the condition causing the fault condition 
has been cleared. When the amplifier is part of a 
multi-amp system, pressing the Reset button on 
the amplifier reporting the fault condition will return 
Figure 4.5 – Main Status Indicators for Stand-Alone Amplifiers
Indicator is lit
Main Status
Indicators
Run
Ready
Standby
Stop
Run
Ready
Standby
Stop
Run
Ready
Standby
Stop
12
Indicator is not lit
State of Operation
Run mode: The amplifier's high-voltage transformers 
are energized and the unit will amplify the input signal. 
Run mode is initiated by: (1) the Enable push button, or 
(2) when the amplifier powers up in Run mode (factory 
default). See the "Advanced Configuration" section for
more information.
Standby mode: Standby mode indicates that the am-
plifier is functioning properly and all Fault Status modes 
are clear, but it is being held in Standby by an external 
condition. As configured from the factory (Run mode on 
startup), the amplifier will enter Standby mode briefly 
after powering up, and then will move automatically into 
Run mode. In Standby mode, the amplifier's low-voltage 
transformer is energized but the high-voltage transformers 
are not.
Stop mode:  When the Stop button on the amplifier front 
panel is pressed, the amplifier will enter Stop mode. The 
amplifier may also enter Stop mode after powering up if 
the amplifier is configured to enter Stop mode on startup. 
In Stop mode, the amplifier's low-voltage transformer is 
energized but the high-voltage transformers are not.
Information subject to change
2105 OPERATOR'S MANUAL – SECTION 4
Figure 4.4 – Main and Fault Status Indicators
all of the amplifiers to Run mode if the condition 
causing the fault condition has been cleared; how-
ever, pressing the Reset button on other amplifiers 
in the system will NOT clear the fault condition.
Refer to the "Fault Status Indicators" section for
information on how to clear fault conditions and
restore amplifier operation.
4.1.4
Main Status Indicators
Four Main Status indicators are located on the
amplifier's front-panel (see Figure 4.4). These 
LEDs monitor the internal conditions of the ampli-
fier and indicate the current state of operation. The 
chart in Figure 4.5 details the operational modes
indicated by the Main Status indicators.
Indicator may be lit
Action Needed to Return to Run
Mode
N/A
If the amplifier remains in Standby mode, 
it is being held in Standby by remote con-
trol through the SIM Interlock I/O connec-
tor. Open the Enable/Standby switch to 
clear this remote Standby condition and
return the amplifier to Run mode. See the 
"Applications" section of this manual
for more information on remote amplifier 
operation.
To release the amplifier from Stop mode, 
press the Enable button.
97-2105006_4-2-15

Advertisement

Table of Contents
loading

Table of Contents