Activating the xBPI Protocol

Note:
Note:

Activating the Service Mode

Purpose
Purpose
Purpose
Purpose
Purpose
The Service mode allows access to an extended menu.
This mode must be activated before you can perform
calibration and adjustment work on the Combics and
on any connected weighing platform.
Activate the xBPI protocol to perform adjustment with the service software
Sartocas software or Psion server without activating the BPI mode
through the SBI/BPI key.
This menu item is accessible in the standard menu.
Setup
Device parameters
Under xBPI a submenu is opened for assigning addresses, but this function is only
available when using xBPI with RS-485.
When the Service mode is active, an „S" is shown on the
right-hand side of the header line in the Setup menu. To
deactivate the Service mode, restart the indicator (turn the
indicator off and back on again).
COM 1 port
Data communications
XBPI-RS232