Tilt Actuator Removal

1. Remove the seating system to expose the
actuator. Disconnect the 6-pin connector by
Pushing down on tab (fi gure 7.9.1)
Caution: The actuator wires are fragile. Use caution when removing. Do not remove the bullet
connector
2. Note the connections leading to the Tilt Actuator,
Separate and label the those wires. Reverse the
procedures to replace it and retest as necessary.
(fi gure 7.9.2).
3. Use two 13 mm wrenches and remove the nuts
at both ends of the actuator. See (fi gure 7.9.3).
which shows one end. Remove both pins from
the tilt actuator.
SUNRISE MEDICAL
Section 7
QM-7 SERIES SERVICE MANUAL
JANUARY 2012
PAGE 7.9
fig. 7.9.1
fig. 7.9.2
fig. 7.9.3