Motor and Gearbox Removal

1. The Motor must be disconnected from power
before removal. For controller access, remove the
controller cover/shroud from the back of the chair
with an M4 Hex Key locating the two attachment
screws. (fi gure.. 7.3.1).
2.
Locate and disconnect main power from
the control module by disconnecting the power
harness from the batteries.
3. Locate and disconnect the appropriate Motor
Lead, from the control module. (M1 , M2 or both)
(fi gure.. 7.3.2).
4. Remove the wheel assembly by using a 5/8"
deep socket wrench to remove the 3 bolts that
hold the wheel and rim in place (fi g. 7.3.3).
5. To separate the Motor from the Motor Bracket,
use a 5mm Hex key to remove the 6 M6 Socket
head Cap screws shown in (fi gure. 7.3.4) .
SUNRISE MEDICAL
Section 7
QM-7 SERIES SERVICE MANUAL
M1
JANUARY 2012
fig. 7.3.1
M2
fig. 7.3.2
fig. 7.3.3
fig. 7.3.4
PAGE 7.3