Table 1- R-Net Error Codes (Cont.) - Quickie QM-7 Series Service Manual

Table of Contents

Advertisement

Table 1- R-NET Error Codes (Cont.)

Trip Text
Over-current
Overtemp. (Acts)
Overtemp. (Lamps)
DIME Error
Memory Error
PM Memory Error
Bad Cable
Bad Settings
Module Error
System Error
SID Detached
User Switch Detached
Gone to Sleep
Charging
SUNRISE MEDICAL
Section 4
Corrective Action
This fault occurs when the limits of an actuator circuit is exceeded.
Perform Tilt Will Not Operate troubleshooting section of this manual to
test the end of travel limit switches.
This error indicates that the Intelligent Seating Module (ISM) has become
excessively warm. Note: An ISM is only required when there are more
than 2 actuators and may not be mounted on this power chair. Allow the
unit to cool. If the error repeats, the actuator may be defective or over
loaded.
Note: This fault can only occur if an Intelligent Seating Module is used on
this power chair and connected to external lights.
This error indicates that one or more of the modules are not compatible.
Remove the last module installed and retest. Recycle the power. If this
error persists, Contact Sunrise Technical Service for assistance.
Recycle the power. Verify that cables are correctly mated. If a module
has been replaced recently, it may be defective. If error persists, contact
Sunrise Technical Service for assistance.
Recycle power. Verify cables are correctly mated. If a module has been
replaced recently, that module may be defective. If error persists, contact
Sunrise Technical Service for assistance.
Inspect and replace the defective cable(s).
Verify that programming agrees with installed equipment. If all settings
are correct, the control module may be defective. If error persists, contact
Sunrise Technical Service for assistance.
Recycle power. Verify that cables are correctly mated. If a module has
been replaced recently, it may be defective. If error persists, contact
Sunrise Technical Service for assistance.
Recycle power. Verify that cables are correctly mated. If a module has
been replaced recently, it may be defective. If error persists, contact
Sunrise Technical Service for assistance.
The Omni has detected that a specialty control has become disconnected.
Recheck all cables. If the error is still present, replace the specialty
control.
Indicates that a user switch has become disconnected.
disconnected switch and reconnect.
The predetermined sleep time has been exceeded due to inactivity by the
user.
This indication is present when the battery charger is connected. There
may also be an error in the control module. If error persists, contact
Sunrise Technical Service for assistance.
QM-7 SERIES SERVICE MANUAL
JANUARY 2012
PAGE 4.5