Press or scroll a or s to move the piston rod
forward until the base plate of the piston rod and
the cartridge plunger are level. The displayed
cartridge volume decreases accordingly.
Re-insert the cartridge into the cartridge
compartment.
– Continue with these instructions when the
cartridge is correctly inserted (the end plate of
the piston rod sits flush against the cartridge
plunger), or
– Repeat this step until the cartridge is correctly
inserted (the end plate of the piston rod sits
flush against the cartridge plunger).
When the cartridge has been inserted correctly
press f. Your insulin pump performs a self
test.
See the section 5.2 "Start-Up procedure" for
further information.
Check that the infusion set luer-lock connector is
still correctly connected to the adapter.
Tighten the infusion set by hand clockwise until
it sits tightly in the adapter.
When the start-up procedure is finished, the
PRIME THE INFUSION SET menu is displayed.
See the section 5.4 "Prime the infusion set" for
further information. Start from the point in those
instructions where the insulin pump screen
matches the menu to the left.
Prepare the pump for use
63