8.1.1 Program an Extended bolus

RUN
screen
stop your
ACCU-CHEK
A8: bolus cancelled
Note Press d to loop between programming the bolus amount
and bolus duration, if necessary.
Extended
bolus
Press d to move to the EXTENDED BOLUS
menu.
Select with f. The BOLUS AMOUNT screen
appears.
Press a to increase or s to decrease the
bolus amount.
The bolus duration of the last Extended bolus
delivered appears, or when using your insulin
pump for the first time, a default value appears.
Press d to move to the BOLUS DURATION.
Press a to increase or s decrease the bolus
duration.
Check the bolus amount and duration
programmed on the display.
ADVANCED user menu
bolus
bolus
amount
duration