Accu-Chek Spirit User Manual page 189

Insulin pump
Hide thumbs Also See for Spirit:
Table of Contents

Advertisement

Bolus
Temporary basal rate
Maximum time for an error
E4: OCCLUSION**
**
Insuman
®
Infusat produced by Aventis Pharma Deutschland GmbH, a member and a brand
of the sanofi-aventis group.
** Determined using measurement methods according to IEC 60601-2-24-1998.
The max. bolus amount per delivery is
25 insulin units. The bolus amount for
the "Quick" Standard bolus is adjustable
in increments of 0.1, 0.2, 0.5, 1.0 and
2.0 units. For the "Scroll" Standard
bolus, the Extended bolus and the
MultiWave bolus, the amount is
adjustable in fixed increments of 0.1
units. The duration of the Extended
bolus and the MultiWave bolus is
adjustable in intervals of 15 minutes
(15 minutes up to 12 hours).
Adjustable in 10 % increments, 0 – 90 %
for decreases, 110 –200 % for increases.
The duration is adjustable in 15 minute
intervals, up to a maximum of 24 hours.
The last programmed duration is given
by default for the next temporary basal
rate change programmed.
Plastic cartridges:
– at med. basal rate 1.0 U/h:
– at min. basal rate 0.1 U/h:
The typical time for an error
was 3.5 hours.
Aventis Insuman Infusat glass
cartridges*:
– at med. basal rate 1.0 U/h:
– at min. basal rate 0.1 U/h:
Technical data
5 h
50 h
10 h
100 h