Before you begin
If the piston rod does not return completely to its base after
the cartridge has been removed, immediately contact your local
Customer Care Area for support.
After the occurrence of an error, your insulin pump will be in
STOP and the insulin delivery is interrupted. In order to maintain
insulin delivery, you must act immediately according to the
instructions given for each error code. Put your insulin pump into
RUN to continue the insulin delivery.
Your insulin pump and its sterile products and accessories
include small component pieces that could pose a choking
hazard to children. Keep sterile products and accessories out of
reach of children.

2.1.5 Warnings and cautions about environmental hazards

Avoid electromagnetic fields of radar or antenna installations,
high-voltage sources, X-Ray sources, MRI, CAT scan or other
sources of electrical current. Do not use your insulin pump in such
areas. Electromagnetic fields may cause malfunction to your insulin
pump. Always stop and remove your insulin pump prior to entering
these areas. In other cases, the insulin delivery may be stopped
immediately and an error E7: ELECTRONIC ERROR occurs.
See the section 10.2.7 "Error E7: ELECTRONIC ERROR" for
further information.
Do not use your insulin pump in hyperbaric chambers and
in hazardous areas of any classification (such as areas where
explosive or flammable gases or vapours could exists) as this
might interfere with insulin delivery and/or lead to harmful
situations.