– During bolus delivery:
• If the Extended bolus delivery has begun, it can be cancelled by
putting your insulin pump into STOP. This cancels the bolus
delivery and an alert A8: BOLUS CANCELLED occurs. Press f
twice to confirm and turn off the alert. See the section
10.1.8 "Alert A8: BOLUS CANCELLED" for further information.
Bolus delivery is interrupted.
The bolus amount delivered up to the cancellation can be
reviewed in the BOLUS HISTORY screen.
See the section 7.8.1 "Review the bolus history" for further
information.
When your insulin pump is in STOP, it does not deliver
any insulin. Put your insulin pump into RUN to continue
the insulin delivery.
Note If a temporary basal rate is active at the same time, it is also
cancelled when your insulin pump is put into STOP. An alert
A8: BOLUS CANCELLED and an alert A6: TBR CANCELLED
occur.
Press f twice to confirm and turn off the first alert. Then
the second alert appears on the display. Press f twice to
confirm and turn off the second alert. Both alerts are
recorded in the alarm history.
See the sections 10.1.6 "Alert A6: TBR CANCELLED"
and 10.1.8 "Alert A8: BOLUS CANCELLED" for further
information.
Ensure that the cancellation was intended and program a new Extended
bolus (and/or temporary basal rate), if necessary.
Caution
ADVANCED user menu