STANDARD user menu
• Put your insulin pump into STOP.
The temporary basal rate delivery is interrupted.
An alert A6: TBR CANCELLED occurs. Press f twice to confirm
and turn off the alert.
See the section 10.1.6 "Alert A6: TBR CANCELLED" for further
information.
When your insulin pump is in STOP, it does not deliver
any insulin. Put your insulin pump into RUN to continue
the insulin delivery.
Note If an Extended bolus or a MultiWave bolus is active at the
same time, it is also cancelled when your insulin pump is put
into STOP. An alert A8: BOLUS CANCELLED and an alert
A6: TBR CANCELLED occur.
Press f twice to confirm and turn off the first alert. Then
the second alert appears on the display. Press f twice to
confirm and turn off the second alert. Both alerts are
recorded in the alarm history.
See the sections 10.1.6 "Alert A6: TBR CANCELLED"
and 10.1.8 "Alert A8: BOLUS CANCELLED" for further
information.
Ensure that the cancellation was intended and program a new temporary
basal rate (and/or an Extended or MultiWave bolus), if necessary.