STANDARD user menu
Note Press d to loop between programming the basal rate
percentage and duration, if necessary.
Note At any screen you have three options to exit:
if you want to confirm and save the changes
– press f. Your insulin pump returns to the RUN screen.
if you want to undo the changes
– wait for your insulin pump to return to the RUN screen
(timeout), or
– press d + a simultaneously (exit feature) to exit
the current function screen.
– set precentage to 100 % and press f.
Note During the TBR delivery, the increase or decrease in percent-
age, the increased or decreased hourly basal rate (hourly
basal rate plus/minus TBR percentage) and the remaining
time are displayed in the RUN screen.
An arrow pointing up symbolizes a temporary basal rate
increase, an arrow pointing down a temporary basal rate
decrease.
Note If an Extended bolus or a MultiWave bolus is active at the
same time, the remaining time and amount of the bolus and
the amount of the increased or decreased hourly basal rate
are displayed in the RUN screen.
104
Press d to move to the TBR DURATION.
Press a or s to correct and set the duration.
Save and exit with f.
The programmed temporary basal rate is
immediately activated.