Alerts and errors

10.2.8 Error E8: POWER INTERRUPT

A power interrupt may occur when you change the battery without
putting your insulin pump into STOP first, or when your insulin pump
was dropped.
If a bolus and/or a temporary basal rate was interrupted by the error,
review the bolus and/or the temporary basal rate history for the delivered
amount and duration.
If an error E8: POWER INTERRUPT occurs during programming, check
your settings.
160
Press f twice to turn off and confirm the error.
Your insulin pump goes into STOP.
Check the time and date and correct,
if necessary.
When finished, put your insulin pump into RUN,
if necessary.
If necessary, program a new bolus and/or
temporary basal rate.