7.5.3 Cancel a "Quick" Standard bolus

– During programming (the bolus amount blinks):
• Press the key you started programming with (s [a]) to reset
the bolus amount to 0.0 units. You hear a melody.
If no new bolus is programmed, your insulin pump returns to the
RUN screen and beeps three times after 5 seconds h.
No bolus is delivered.
– During confirmation (when the beeps and vibrations occur), or during
the start delay (
• Press s or a. You hear a melody. Your insulin pump returns
to the RUN screen. An alert A8: BOLUS CANCELLED occurs.
Press f twice to confirm and turn off the alert. See the section
10.1.8 "Alert A8: BOLUS CANCELLED" for further information.
No bolus is delivered.
– During bolus delivery (countdown of bolus amount):
• Press and hold the s or a key for 3 seconds until you hear a
melody.
An alert A8: BOLUS CANCELLED occurs. Press f twice to
confirm and turn off the alert. See the section 10.1.8 "Alert A8:
BOLUS CANCELLED" for further information.
Bolus delivery is interrupted.
The actual bolus amount delivered prior to the cancellation
can be reviewed in the bolus history. Please refer to the section
7.8.1 "Review the bolus history" for further information.
Ensure that the cancellation was intended and program a new Standard
bolus, if necessary.