Alerts and errors

10.1.2 Alert A2: BATTERY LOW

Your insulin pump checks the battery voltage before each insulin delivery
(at least every 3 minutes). If it falls below a certain voltage value, the alert
A2: BATTERY LOW occurs. Be prepared for a battery change as quickly as
possible. The battery low symbol (
STOP screen as a reminder until you insert a new battery.

10.1.3 Alert A3: REVIEW TIME AND DATE

If the battery is removed for more than 1 hour, it may be necessary to
enter the time and date again. Other Accu-Chek Spirit settings and the
event memory are not affected and are preserved, regardless of battery
voltage and time your insulin pump has been without a battery.
148
) appears in the RUN screen or
Press f twice to turn off and confirm the alert.
Replace the battery within the next hours
Put your insulin pump into RUN, if necessary.
Press f twice to turn off and confirm the alert.
Review the time and date and correct them,
if necessary.
Put your insulin pump into RUN, if necessary.