Prepare the pump for use
To avoid leakage, properly tighten the infusion set to
the adapter. Turn the luer-lock connector only as far as
to the stop. Do not turn it any further and do not use
any auxiliary tools as this may crack the infusion set
luer-lock connector and result in leakage.
Note Your insulin pump cannot detect infusion set leakage. You
must inspect all parts of the infusion set at least every three
hours during the day and before you go to sleep. Should you
detect any loss of insulin and all parts are correctly
tightened, immediately replace the leaking component.
Because insulin delivery has been interrupted, check your
blood glucose level and take appropriate actions according to
your doctor's or healthcare team's instructions.

5.3.2 Insert the cartridge

Do not carry out the CHANGE THE CARTRIDGE function
if the cartridge compartment is not completely dry.
During the rewinding of the piston rod liquid might enter
your insulin pump and could lead to malfunction.
Should the piston rod not return completely, contact your local Customer
Care Area for support.
60
Caution
Caution