Accu-Chek Spirit User Manual page 194

Insulin pump
Hide thumbs Also See for Spirit:
Table of Contents

Advertisement

Technical Data
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
192
10 V/m
Portable and mobile RF
communications equipment
80 MHz to 2.5
should be used no closer to any
GHz
part of the Accu-Chek Spirit,
including cables, than the rec-
ommended separation distance
(IEC 60601-2-24)
calculated from the equation
applicable to the frequency of
the transmitter.
Recommended separation
distance:
d = 0.4
d = 0.7
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
(a), should be less than the
compliance level in each
frequency range (b).
Interference may occur in the
vicinity of equipment marked
with the following symbol:
80 MHz
P
to 800 MHz
800 MHz
P
to 2.5 GHz

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents