7.7.4 Automatic off

Discuss the use of the automatic off function with your
doctor or healthcare team.
Automatic off is a safety feature which stops all insulin delivery by
triggering an error E3: AUTOMATIC OFF if no keys are pressed within a
programmed time period in RUN.
The automatic off can either be set to OFF or programmed up to 24 hours
in 1 hour intervals in the SETUP MENU STANDARD.
Press d to move to the ALARM SIGNALS
screen. The active alarm signals are displayed.
Press a or s to set the alarm signals as
desired.
Save and exit with f.
Caution
Press d to move to the SETUP MENU
STANDARD.
Select with f.
Press d to move to the AUTOMATIC OFF
screen. The active status is displayed.
Press or scroll a to increase or s to
decrease the duration for the automatic off in 1
hour intervals. If you would like to set the auto-
matic off function to OFF, press or scroll s
until 0 hours and OFF appear on the display.
Save and exit with f.