Calibration Functions on Standard Balances

Select the appropriate menu code to define the access status for each
of the calibration functions, which are activated by holding down the t key for
a few seconds.
If the menu access switch is unlocked ("accessible" status indicated by -C- after
you have accessed the balance operating menu), the "external calibration"
function will be accessible even though you have set the menu code 1 9 2 for
"access denied."
External calibration
Accessible
Access denied
Internal calibration
Accessible
Access denied
Calibration test
Accessible
Access denied
External Linearization
The linearization weights to be loaded are displayed on the balance one
after the other in increasing order. The balance must be re-calibrated after
external linearization.
Important Note:
If you use linearization weights that differ from the values displayed, this will
cause errors.
External linearization
Accessible
Access denied
Internal linearization
Accessible
Access denied
* = factory setting
2–14
Code
1
9
1*
1
9
2
Code
1 10
1*
1 10
2
Code
1 11
1*
1 11
2
Code
1 12
1
1 12
2*
Code
1 13
1*
1 13
2