Weighing in Percent

Weighing in Percent
Symbol displayed: %
Key identified by: STO
This application program enables you to obtain weight readouts in percent which
are based on a reference weight "Wxx%."
You can store a weight as a reference percentage "pRef" using the F key,
identified by "STO REF," in the following ways:
– Place a reference sample on the pan and press the function key identified
by "STO REF" to have the reference weight for 100% calculated from the sample
weight on the pan "Wxx%" and from the reference percentage "pRef" (shown
on the application display).
– Enter a reference percentage "pRef" using the numeric keys and store it with the
function key identified by "STO REF" to have the reference weight for 100%
calculated based on the percentage entered and on a weight on the pan.
To change the standard percentage (value shown on the application display),
either access the parameter mode or enter a different number using the
numeric keys (see page 3–49). You can read the reference percentage in the
application display as long as you have not yet stored the reference weight.
Use w to toggle between percent and weight readouts.
Important Note for Verified Balances Approved for Use as
Legal Measuring Instruments in the EU:*
You may not convert a weight into a percent when the balance is being used
as a legal measuring instrument. The display symbol "R2" flashes to identify non-
verified display readouts (not legal for trade).
* including the Signatories of the Agreement on the European Economic Area
REF - F