Density

)
1
Density
Symbol displayed:
Key identified by: STO - F
You can determine the density of solid, liquid and powdery samples according
to one of two methods:
– Buoyancy method
– Specific gravity bottle method
The Achimedean principle is applied to determine the volume of a particular
sample; this volume is necessary for calculating the density.
Methods
Buoyancy Method
Specific Gravity
Bottle Method
Select the method of your choice by setting the corresponding menu code from
the list below.
Density determination method
Buoyancy
Specific gravity bottle
Important Note Concerning Balances Used as Legal Measuring Instruments
in the EU**
The displayed result of the density determination procedure is not a verified
value. The display symbol "R2" flashes to identify non-verified values
(not legal for trade).
) = not for MC 5 (-0CE), SC 2 (-0CE) and MC 21 S microbalances
1
**
= factory setting
** = including for Signatories of the Agreement in the European Economic Area
3–34
To measure the density inside the weighing chamber
(toploading configuration), use the YDK 01 (-0D).
The method using a specific gravity bottle is suitable
for determining the density (specific gravity) of pasty and
powdery samples.
Code 2 1 7
Code
2
6
1*
2
6
2

Hide quick links: