Internal Fuses; F Use Sizes And Fuse Replacement - Carbolite MRF 16 Installation & Operating Instructions Manual

Table of Contents

Advertisement

MRF 16/22
13.5
Fuse sizes and fuse replacement
Access to fuses for the heating elements and supply to the microwave circuit is by removal
of the left side cover [See figure 13.31].
Power
Phase
Supply
Fuses
F1, F2, F3
208 -
3 x 40A
240V
3 Phase
14 x 51mm
Delta
Type GG
380 -
F1, F2, F3
415V
3 x 25A
3 Phase
plus
10 x 38mm
Neutral
Type GG
(STAR)
To access fuses F1 to F13 isolate the furnace from the electrical supply and remove the left
side cover. [See figure 13.1].
To access fuses F14 to F16 isolate the furnace from the power electrical power supply and
remove the right side access panel followed by the removal of the High Voltage Circuits
cover. Note that the High Voltage Circuit could have energy stored in the capacitors, so
ensure that they are discharged before replacing any fuses.
44

Internal Fuses

Aux Fuse
Line Fuses
F4, F5
F6, F7
2 x 5A
2 x 25A
5 x 20mm
10 x 38mm
Type F
Type GG
F4, F5
2 x 5A
N/A
5 x 20mm
Type F
Fig 14.2 Internal fuses
Thyristor
Microwave
Fuses
S.C.
F8, F9,
F14
F10, F11
4 x 30A
1 x 20A
10 x 38mm
10 x 38mm
Protistor
Type GG
F8
F14
1 X 30A
1 x 20A
10 x 38mm
10 x 38mm
Protistor
Type GG
HT Fuses
F15, F16
2 x HVF10
F15, F16
2 x HVF10
MF 55-GB 1.21

Advertisement

Table of Contents
loading

Table of Contents