R Unning The Current Program - Carbolite MRF 16 Installation & Operating Instructions Manual

Table of Contents

Advertisement

Segment Type
Press scroll
until the display shows "Segment Type". Press up
Rate, Time, Dwell, Step, Wait, GoBack, Call or End.
Target Setpoint (visible only for Rate, Time and Step segments)
Press scroll
until the display shows "Target SP". Press up
Ramp Rate (visible only for Rate segments)
Press scroll
until the display shows "Ramp Rate". Press up
of degrees per Ramp Unit as set above.
Duration (visible only for Time and Dwell segments)
Press scroll
until the display shows "Duration". Press up
display shows "hrs:mins:sec".
Holdback Type (visible only for Rate, Time and Step segments)
Press scroll
until the display shows "Holdback Type". If required, press up
select Off, Low, High or Band. [See section 8.9].
Wait For (visible only for Wait segments)
This segment function is not used.
GoBack Seg (visible only for GoBack segments)
The program will go back once to the beginning of a chosen segment and repeat through to
the end. Select a segment number for the program to go back to the start of.
Call Program (visible only for Call segments)
This segment function is available on the P10, P25 and P50. Select another program number
to run as an intermediate program.
Call Cycles (visible only for Call segments)
Set the number of time to repeat the intermediate (called) program.
Micro % (this can be set in Rate, Time and Dwell segments)
Set the MW power level for that segment. Read the following notes about Event Outs.
Event Outs (switches MW power on/off in segments)
If both squares are empty, MW power will not come on in that segment.
If both squares are full, MW power will come on in the segment at the power level set on
as in 'Micro %' above.
If the first square is highlighted and the second square is blank, MW power will come on in
the segment at the MW power level set on the MW control page, it can be changed manually
while a program is running.
If the first square is blank and the second square is highlighted, MW power will not come
on in the segment.
End Type (visible only for End segments)
Press scroll
until the display shows "End Type". Press up
Dwell or SafeOp.
To reduce the number of segments in a program, change the last required segment to an End
type. Press
for OK or
To cancel all segments in a program, change the first segment to an End type.
8.3
Running the current program
The current program can be started by pressing RUN/HOLD twice.
MF 55-GB 1.21
to cancel.
or down
to select
or down
to set a value.
or down
to set the number
or down
to set a value. The
or down
or down
to select Reset,
MRF 16/22
to
25

Advertisement

Table of Contents
loading

Table of Contents