Press scroll
minute (500°C/hr).
Press scroll
Press scroll
Press scroll
or down
to fill the first square in
Press scroll
The second square will flash. Press up
square in
.
For segments 2,3,4 and 5 enter the following parameters and values:
Segment 2: Dwell, Duration 30:00.0 minutes, Micro 100%, Event outs
Segment 3: Rate, Target SP 1400°C, 5.0°C/min (300°C/hr), Micro 100%, Event outs
Segment 4: Dwell, Duration 30:00.0 minutes, Micro 100%, Event outs
Finish the program with an End segment:
Segment 5: End, Reset.
To run the program:
Press page
and scroll
Press RUN/HOLD twice
To pause (hold) the program:
Press RUN/HOLD once
To reset the program:
Press RUN/HOLD once to pause (hold) it
Press RUN/HOLD for 3 seconds.
MF 55-GB 1.21
Ramp Rate is displayed. Press up
Holdback Type is displayed. Default 'Off'
Micro% is displayed. Press up
Event outs
is displayed. The first square will be flashing. Press up
at the same time return to the home page.
or down
or down
.
to set to '8.3' degrees per
to set to '50%'
or down
to fill the second
.
.
MRF 16/22
.