Carbolite MTF 15 Installation, Operation & Maintenance Instructions Manual

900c tube furnace

Advertisement

Quick Links

Installation, Operation &
Maintenance Instructions
This manual is for the guidance of operators of the above Carbolite product and
should be read before the furnace is connected to the electricity supply.
section
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
This manual should supply all the information required for safe and trouble-
free furnace operation. Information on controller operation is included.
1.0
900°C Tube Furnace
type MTF 9/15
C
ONTENTS
MF09
page
2
3
4
5
10
11
13
14
15
16

Advertisement

Table of Contents
loading

Summary of Contents for Carbolite MTF 15

  • Page 1: Table Of Contents

    Installation, Operation & Maintenance Instructions 900°C Tube Furnace type MTF 9/15 This manual is for the guidance of operators of the above Carbolite product and should be read before the furnace is connected to the electricity supply. ONTENTS section page Symbols &...
  • Page 2: Symbols & Warnings

    MTF 9/15 & W YMBOLS ARNINGS Switches and Lights Supply Light: when the furnace is connected to the electrical supply the light in the adjacent switch glows Heat Light: the adjacent light glows or flashes to indicate that power is being supplied to the elements Warning Symbols DANGER of electrical shock–...
  • Page 3: Installation

    MTF 9/15 NSTALLATION Unpacking & Setting Up Unpack the furnace carefully and remove any packing from inside the tube. Place the furnace in a well ventilated area, away from other sources of heat, on a surface which is resistant to accidental spillage of hot materials. Do not use the furnace on an inflammable surface. Ensure that there is free space around the furnace.
  • Page 4: Operating Instructions

    (grounded). Avoid burns. Carbolite can supply tongs, face masks, and heat resistant gloves. Before you remove a hot object from the furnace make sure you have a safe place to put it down.
  • Page 5: Controller Operation

    MTF 9/15 ONTROLLER PERATION Eurotherm 2132 When switched on, the controller lights up, goes through a short test routine, and then displays the measured temperature and starts to control. The output light glows or flashes as heating occurs. The Page key allows access to parameter lists within the controller;...
  • Page 6 MTF 9/15 Altering the Ramp Rate SPrr It is possible to limit the rate of heating by setting a ramp rate. Press Scroll until the legend (SetPoint ramp rate) is displayed. Use Down or Up to display and adjust the value. The ramp rate sets the maximum rate of heating or cooling in degrees per minute.
  • Page 7 MTF 9/15 Operation With the Timer The 2132 can be used as a process timer allowing timed heating or timed delay, according to the options in the table. There are 5 timer modes, but 2 of them are affected by whether the setpoint ramp rate feature is being used, making 7 entries in the table.
  • Page 8 MTF 9/15 Setting the Timer Mode .1 to tm.OP OPt.5 Scroll to ; use to view and alter the mode. The mode shows as It is not possible to alter the mode while the timer is running; if the mode cannot be altered, scroll StAt to the parameter and set its value to...
  • Page 9 MTF 9/15 4.11 Stopping the Timer StAt To stop the timer at any time while it is running, change the parameter to . This is the same as reducing to zero. The controller then acts as though at has reached the end of the time period.
  • Page 10: Maintenance

    To contact Thermal Engineering Services or Carbolite see the back page of this manual. Recommended Spares Kits Carbolite can supply individual spares, or a kit of the items most likely to be required. Ordering a kit in advance can save time in the event of a breakdown.
  • Page 11: Repairs & Replacements

    MTF 9/15 & R EPAIRS EPLACEMENTS Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnected from the supply before repair work is carried out. Safety Warning - Refractory Fibrous Insulation This furnace contains refractory fibres in its thermal insulation. These materials may be in the form of fibre blanket or felt, vacuum formed board or shapes, mineral wool slab or loose fill fibre.
  • Page 12 MTF 9/15 Upon reassembly ensure that the element wire is deeply embedded in the grooves in the ends of the insulation and held away from the foil by insulation beads. The foil should be neatly and firmly wrapped so that it positively locates the beads. When refixing the wrapped element position the beads just above edge of the metal channel to maximise the clearance between the outer case and the wires.
  • Page 13: Fault Analysis

    MTF 9/15 AULT NALYSIS A. Furnace Does Not Heat Up The HEAT light The heating Check also that the SSR is working element has failed correctly is ON The HEAT light The controller The thermocouple has broken or has a wiring fault is OFF shows a very high...
  • Page 14: Circuit Diagram

    MTF 9/15 IRCUIT IAGRAM Circuit Diagram grey heating controller filter pink element thermocouple Blue Blue Green/Yellow MF09...
  • Page 15: Fuses & Power Control

    MTF 9/15 & P USES OWER ONTROL Fuses Access to internal fuses is by removal of the outer case. Fuses: 2A 20mm x 5mm type F. Power Control In this model the same heating element is used for 110-120V models and 220-240V models. Electronic power limiting in the 2132 controller is used to reduce the power in the 220-240V model.
  • Page 16: Specifications

    Thank you for reading this data sheet. Please contact us for a quotation: Head Office (UK) Switzerland Office P.O. Box 2124 Zugerstrasse 46 Chelmsford 6314 Unterägeri Essex CM1 3UP England Switzerland Tel: +44 (0)1245 600 560 Tel: +41 41 541 5877 Fax: +44 (0)1245 600 030 Fax: +41 41 541 5878 Email:...