FUSES & POWER SETTINGS
8.1
Fuses -
For wiring diagram information on F1 and F2 refer to Figure 7-1 page 22, for fuse values refer to Table
8-2 and Table 8-3.
Table 8-1. Fuse type information.
Internal
F1
supply fuses
Auxiliary
F2
circuit fuses
Customer
fuses
8.2
Table of internal fuse and external supply ratings.
All furnace models are single phase. Most models have internal fusing as in the Table 8-2 and
Table 8-3 below. Customers should check the rated amps in the tables below and make sure that
their electrical supply is rated accordingly (or at least the next fuse size up from the rated amps
table). Customers should also fit external fusing if their particular model of furnace does not have
internal fuses. The function of the customer's (external) fuse is to protect the customer's wiring.
All electrical connections other than where it is just a simple plug into an existing correctly rated
electrical supply must be carried out by an electrically competent person with the correct
certification to meet local legislation.
Table 8-2. Fuse ratings for 100 to 110V Supply Voltages
EHA/EVA/EST/EVT 12/150
Single
EHA/EVA/EST/EVT 12/300
Zone
EHA/EVA/EST/EVT 12/450
EHA/EVA/EST/EVT 12/600
Three
EHC/EVC/EZS/EVZ 12/450
EHC/EVC/EZS/EVZ 12/600
Zone
Table 8-3. Fuse ratings for 200 to 240V Supply Voltages
EHA/EVA/EST/EVT 12/150
Single
EHA/EVA/EST/EVT 12/300
Zone
EHA/EVA/EST/EVT 12/450
EHA/EVA/EST/EVT 12/600
Three
EHC/EVC/EZS/EVZ 12/450
Zone
EHC/EVC/EZS/EVZ 12/600
Fitted to all E range models.
Each circuit has 2 fuses, in most cases fitted to the
EMC filter board.
Fitted to all E range models.
In most cases fitted to the EMC filter board.
Required to protect customer supply circuit and
lead.
Model
Model
MF65-1.01
Power
Internal
rating
supply
(Watts)
fuses F1
2x10A
750
2x16A
1480
2x20A
1960
2x25A
2480
2x20A
1960
2x25A
2480
Power
Internal
rating
supply
(Watts)
fuses F1
2x4A
750
2x10A
1480
2x10A
1960
2x12A
2480
2x10A
1960
2x12A
2480
32mm x 10mm type F
2 Amps
On board: 20mm x 5mm type F
Other: 32mm x 6mm type F
See rating label for amperage; see
Table 8-2 and Table 8-3for
minimum supply and fuse ratings.
Furnace
Internal
rated amps
instrument
1-phase
fuses F2
100-110V
2x2A
7A
2x2A
14.5A
2x2A
19A
2x2A
24A
2x2A
19A
2x2A
24A
Furnace
Internal
rated amps
instrument
1-phase
fuses F2
200-240V
2x2A
3.5A
2x2A
7.1A
2x2A
9.5A
2x2A
12A
2x2A
9.5A
2x2A
12A